Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Yokogawa HR2300: (1 year) CAL VER /5520 DATE: 15-Sep-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.1 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 26 NUMBER OF LINES: 123 CONFIGURATION: Fluke 5520A ============================================================================= # # Compatibility: # 5500/CAL 5.0 or MET/CAL 5.0 or later. # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5520A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- J F W 1.002 ASK+ K 1.003 HEAD PRELIMINARY INSTRUCTIONS 1.004 DISP Connect UUT to AC power and turn unit ON. 1.004 DISP 1.004 DISP Warm-up time: 30min - REQUIRED 1.004 DISP Temperature to 23 +/-2 degC 1.004 DISP Humidity: 55 +/-10% 1.005 DISP [32] WARNING 1.005 DISP HIGH VOLTAGE is used or exposed during the performance 1.005 DISP of this calibration. DEATH ON CONTACT may result if 1.005 DISP personnel fail to observe safety precautions. 1.006 HEAD {MEASUREMENT} 1.007 DISP Connect the 5520A to the UUT as follows: 1.007 DISP [32] 5520A NORMAL HI to UUT +INPUT TERMINAL 1.007 DISP [32] 5520A NORMAL LO to UUT -INPUT TERMINAL 1.008 DISP Set FUNCTION to VOLTS. 1.008 DISP Set CHANNEL # to 1. 1.009 HEAD MEASUREMENT: {20mV Range} 1.010 5520 20 0.000mV 0.005U 2W 2.001 5520 20 20.000mV 0.05% 0.005U 2W 3.001 HEAD MEASUREMENT: {60mV Range} 3.002 5520 60 0.00mV 0.02U 2W 4.001 5520 60 60.00mV 0.05% 0.02U 2W 5.001 HEAD MEASUREMENT: {200mV Range} 5.002 5520 200 0.00mV 0.02U 2W 6.001 5520 200 200.00mV 0.05% 0.02U 2W 7.001 HEAD MEASUREMENT: {2V Range} 7.002 5520 2 0.0000V 0.0002U 2W 8.001 5520 2 2.0000V 0.05% 0.0002U 2W 9.001 HEAD MEASUREMENT: {6V Range} 9.002 5520 6 0.000V 0.002U 2W 10.001 5520 6 6.000V 0.05% 0.002U 2W 11.001 HEAD MEASUREMENT: {20V Range} 11.002 5520 20 0.000V 0.02U 2W 12.001 5520 20 20.000V 0.05% 0.02U 2W 13.001 HEAD MEASUREMENT: {50V Range} 13.002 5520 50 0.00V 0.02U 2W 14.001 5520 50 50.00V 0.05% 0.02U 2W 15.001 HEAD {RECORDING} 15.002 RNG 2V 15.003 DISP The following tests verify the recording PEN reflects 15.003 DISP the applied voltages. Check using a glass scale or 15.003 DISP ruler capable of .25mm measurements. 15.004 DISP Channel 1 ZONE must be set to 0~250mm by performing the 15.004 DISP following: 15.004 DISP Press AUX, NEXT, F3, ENTRY. 15.005 DISP Press 0, ENTRY, 250, ENTRY. 15.006 DISP Press the START/STOP key to start the chart recorder. 15.007 5520 -2.0000V S 2W 15.008 EVAL Is the {PEN MOVEMENT 0mm +/- .25mm}? 16.001 5520 -1.0000V S 2W 16.002 EVAL Is the {PEN MOVEMENT 62.5mm +/- .25mm}? 17.001 5520 0.0000V S 2W 17.002 EVAL Is the {PEN MOVEMENT 125mm +/- .25mm}? 18.001 5520 1.0000V S 2W 18.002 EVAL Is the {PEN MOVEMENT 187.5mm +/- .25mm}? 19.001 5520 2.0000V S 2W 19.002 EVAL Is the {PEN MOVEMENT 250mm +/- .25mm}? 20.001 HEAD {RTD} 20.002 DISP Connect the 5520A to the UUT Channel 1 as follows: 20.002 DISP [32] 5520A NORMAL HI to A INPUT TERMINAL 20.002 DISP [32] 5520A NORMAL LO to B INPUT TERMINAL 20.002 DISP [32] 5520A NORMAL LO to b INPUT TERMINAL 20.002 DISP All leads should be equal length and of the same type 20.002 DISP wire and connectors 20.003 DISP Set FUNCTION to RTD. 20.003 DISP Set RANGE to PT4. 20.004 5520 -200.0degC 0.05% 0.3U R1 2W 21.001 5520 0.0degC 0.05% 0.3U R1 2W 22.001 5520 500.0degC 0.05% 0.3U R1 2W 23.001 HEAD {REFERENCE JUNCTION COMPENSATION} 23.002 DISP Short circuit the High and Low terminals using 23.002 DISP standard copper wire. 23.003 DISP Set RANGE to the thermocouple type you use. 23.003 DISP Any TC is acceptable. 23.004 OPBR Is the selected thermocouple type R,S,B or W? 23.005 JMPF 24.002 23.006 EVAL Does the UUT indicate {AMBIENT temperature +/- 1degC}? 24.001 JMP 25.001 24.002 EVAL Does the UUT indicate {AMBIENT temperature +/-.5degC}? 25.001 HEAD {CHART SPEED} 25.002 DISP Set CHART SPEED to 1000mm/h. 25.002 DISP 25.002 DISP Start feeding chart by pressing START/STOP key. 25.002 DISP After one hour, stop feeding chart by pressing 25.002 DISP START/STOP key. 25.003 EVAL Does {chart measure between 999mm and 1001mm?} 26.001 END